Sirona ORTHOPHOS SL 2D Service Manual page 296

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

8 Service routines
8.1 Perform service routines via the control panel
Display
Error
01
Normal state
02
Plug X1000 is not plugged in on
board DX1
03
Persistent ADC values
04
Interruption of the 5 V line
05
Interruption of V
06
Interruption of V
07
Interruption of V
08
Interruption of "Connect" line
09
Defective choke, pin 3 detached
10
Defective choke, pin 1 detached
11
Defective clock generator
12
Resistor R11 not soldered
13
Plug on occlusal bite block not
inserted
14
Composite error
15
Calculated angle too small
16
Calculated angle too large
17
Error on ADC channel 2
296
8.1.15.7.1
Codes of possible error modes
Troubleshooting measures
No action
Check plug for occlusal bite block on board DX1.
Check cabling and plug connections on board DX1 and occlusal
bite block electronics.
line
ref-
line
out-
line
GND
Replace occlusal bite block electronics (board DX51).
Check plug for occlusal bite block on occlusal bite block
electronics (board DX51).
Check cabling and plug connections on board DX1 and occlusal
bite block electronics.
Perform a readjustment via service routine S018.9.
Replace occlusal bite block electronics (board DX51); check
cabling and plug connections on board DX1 and occlusal bite
block electronics.
You can draw conclusions concerning the correctness of the angle
acquisition based on the plausibility of these displays. A plausible
constellation (reference unit) would be:
+22° → AD = 1273
-17° → AD = 3124
The settings may differ unit specifically.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016