9.6.2

Replacing the control panel

1. Remove the user interface [ → 349].
B
D
2. Swing the folding mechanism for the Easypad (A) all the way up. This
makes it easier to remove the cover.
3. Carefully remove the adhesive strip from the shield plate (B) and
grounding strap (C).
4. Press into slit (D) of the housing cover with a screwdriver (do not pry!)
and remove the cover.
5. Detach cables L9 and L10 from the strain reliefs.
6. Rotate the control panel to the center position and loosen the four
screws (E).
7. Pull the cables out of the control panel and remove the panel.
8. Pull the cables L9 and L10 into the new control panel and install it with
the four screws (E).
9. Lay the cables and reattach the strain reliefs (see the chapter Laying
cables when replacing the control panel [ → 352]).
10. Guide the grounding strap (C) of the cover into the new control panel.
11. Clean the adhesive surface and affix the grounding strap (C) with an
adhesive strip to the shield plate (B).
12. Reattach the covers.
13. Install the user interface [ → 349].
9.6 Control panel
A
C