Technical Data; Unit Data - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

3 Unit description
3.1 Hardware
40
3.1.9

Technical data

Unit data

3.1.9.1
Model designation:
Nominal voltage:
Permissible fluctuation:
Permissible drop under load: 10%
Rated current:
Nominal power output:
Nominal frequency:
Mains resistance:
Main building fuse:
Power consumption:
Power output of tube
assembly:
Tube voltage:
Tube current:
Maximum setting range:
High-voltage waveform:
High voltage generation
frequency:
Program duration:
Exposure time:
Image acquisition scale:
Exposure time for a
cephalometric image:
Image acquisition scale for a
cephalometric image:
Total filtration of X-ray tube
assembly:
Focal spot size as specified
in IEC 60336, measured in
the central X-ray beam:
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
ORTHOPHOS XG 3D/Ceph
200 – 240 V
± 10%
12A
2 kW at 90 kV/12 mA with any radiation
time
50 Hz / 60 Hz
max. 0.8 ohms
25 A slow-blow (16 A for single line)
2 kVA
69 kV / 16 mA = 1104 W with any
radiation time
60 – 90 kV (for 90 kV max. 12 mA)
3 – 16 mA (for 16 mA max. 69 kV)
60 kV / 3 mA to 90 kV / 12 mA
High-frequency multipulse
Residual ripple ≤ 4 kV
40 – 120 kHz
See "Program values".
See "Program values".
For P1, normal dental arch (slice center)
approx. 1:1.19, i.e. the acquired image is
magnified by approx. 19% on average
compared to reality.
14.9 s max.
approx. 1:1.1, i.e. the acquired image is
magnified by approx. 10% on average
compared to reality.
> 2.5 mm AI / 90 IEC 60522
0.3 mm Cu for volume exposures
0.5 mm
D3632.076.01.03.02 09.2016