9 Repair
9.2 Height adjustment motor (M1_4)/spindle
move the slide
C
F
E
334
9.2.1.1.2
Moving the slide without the "height adjustment" service kit,
REF. 62 57 518
1. Loosen the two screws (C) and remove the cover (D).
C
D
2. Loosen the 1st of the two screws (E) on the spindle holder (F).
3. Attach the socket wrench (SW19) to the spindle. Hold it firmly in place
while you unscrew the 2nd of the two screws (E).
CAUTION! If the socket wrench has to be reset, secure the spindle
holder against turning, e.g. by using a screw.
4. Rotate the spindle holder using a socket wrench (SW19) to move the
slide to the required height.
Clockwise rotation of spindle = slide moves up
Counterclockwise rotation of spindle = slide moves down
5. After reaching the desired target position, secure the position again
using the two screws (E).
CAUTION! Before replacing the height adjustment motor, the slide
must be secured in this position.