7.2.1.1.2
Manual adjustment: Panoramic sensor
1. Start the adjustment procedure as described in the chapter entitled
"Automatic adjustment: Panoramic sensor (standard) [ → 182]".
2. After taking the exposure, measure the distances S1, S2 and S3 with
the toolbar measuring ruler [ → 166].
To measure S1, S2, and S3, estimate where the horizontal center of
the displayed needles is located. Take the measurement in the lower
area of the needles if possible, since they may have become slightly
bent after repeated use.
Tip: For better measurement, you can also edit [ → 166] the exposure
using the toolbar functions.
7 Unit adjustment and calibration
7.2 Adjustment and calibration via the calibration menu