7 Unit adjustment and calibration
7.2 Adjustment and calibration via the calibration menu
212
Adjustment of the ceph secondary diaphragm (QuickShot)
7.2.1.7
7.2.1.7.1
Automatic adjustment: Ceph secondary diaphragm (QuickShot)
(standard)
NOTICE
Phantom for Ceph exposure
Risk of damage to unit
It is essential that the needle phantom is removed from the bite block
holder on the unit before a ceph exposure is taken; otherwise, the
sensor could collide with the phantom.
The pan needle phantom is
unit.
The ceph test phantom does
unit.
1. In the structure tree, under
secondary diaphragm (Quickshot)"
The
menu is displayed in the action area.
2. Establish [ → 175] receptivity.
not
in the pan bite block holder on the
not
connect to the cover of the sensor
"Ceph"
"Adjustment of Ceph secondary diaphragm (Quickshot)"
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
"Adjustment of Ceph
, click on the
element (S010.8).
64 95 258 D3632
D3632.076.01.03.02 09.2016