5 Messages
5.4 List of error messages
Error code
Description
E4 13 14
Actuator 2; position counter error.
Error code
Description
E6 13 15
Actuator 1 is not ready for operation This error is a sequential fault.
Error code
Description
E6 13 17
Error when activating actuator 1.
Error code
Description
E4 13 21
Ring motor has not reached home
position.
100
Actions required
E4 13 14
● Check the actuator mechanisms manually
for smooth and easy running.
In case of stiffness ...
● Manually reset the rotary ring to the pan
home position and run a test cycle via P1.
If the error occurs again ...
● Check light barrier V1_2 (X803) with
service routine S015.5, replace light
barrier if necessary.
● Run service routine S015.5 (free travel) to
check electronic activation and jerk-free
running, replace actuators and/or spindles
and board DX1 if necessary.
Actions required
E6 13 15
● Restart the unit:
1. Switch off the unit.
2. Wait 1 minute.
3. Switch unit on.
4. Repeat procedure and observe causal
error messages.
Actions required
E6 13 17
● Restart the unit:
1. Switch off the unit.
2. Wait 1 minute.
3. Switch on the unit.
4. Repeat procedure.
● Please contact the Sirona Customer
Service Center (CSC) to find out whether
a bug fix by means of a software update is
possible and perform such an update if
necessary.
Actions required
E4 13 21
● Check the ring drive mechanism manually
for smooth and easy running, replace the
ring motor or mechanism if necessary.
● Check light barrier V1_3 (X804), replace if
necessary.
● Replace board DX1.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
see
S. [ → 276],
S. [ → 330]
see
see
S. [ → 68]
see
S. [ → 414],
S. [ → 342]
S. [ → 151],
S. [ → 393]
S. [ → 396]
64 95 258 D3632
D3632.076.01.03.02 09.2016

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orthophos sl 3dOrthophos sl 3d ceph

Table of Contents