Location 83: Sensor Unit - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

5 Messages
5.4 List of error messages
Error code
Description
E7 83 01
Malfunction when starting the
sensor unit.
Communication problems following
unit start.
Error code
Description
E6 83 03
Communication fault for sensor unit. ● Check plug connections of fiber-optic
Error code
Description
E6 83 07
Communication fault for sensor unit. ● Check plug connections of fiber-optic
Error code
Description
E7 83 12
Sensor unit has been taken out of
operation due to malfunction.
Error code
Description
E6 83 15
The DHCP server does not provide
an IP address.
Error code
Description
E6 83 16
The network configuration has been
reset.
134
5.4.15

Location 83: Sensor unit

Actions required
E7 83 01
● Check plug connections of fiber-optic
Actions required
E6 83 03
Actions required
E6 83 07
Actions required
E7 83 12
● Check plug connections of fiber-optic
Actions required
E6 83 15
● Check plug connections of fiber-optic
Actions required
E6 83 16
● Perform network configuration for the unit.
If the error occurs repeatedly ...
● Replace sensor unit.
cables to boards DX83 (X800) and DX1
(J901).
cables to boards DX83 (X800) and DX1
(J901).
cables to boards DX83 (X800) and DX1
(J901).
cables to boards DX83 (X800) and DX1
(J901).
cables to boards DX83 (X700) and media
converters.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
see
S. [ → 30]
see
S. [ → 30]
see
S. [ → 30]
see
S. [ → 30]
see
S. [ → 30]
see
S. [ → 378]
64 95 258 D3632
D3632.076.01.03.02 09.2016