7 Unit adjustment and calibration
7.2 Adjustment and calibration via the calibration menu
192
2. Once you have taken the exposure, measure distances Sx and Sy
with the toolbar measuring ruler [ → 166].
Tip: For better measurement, you can also edit [ → 166] the exposure
using the toolbar functions.
3. Overwrite the adjustment values for Sx and Sy in the text boxes of the
"Diaphragm adjustment"
For information about the shifting direction (input of +/- sign in the
menu), see the chapter entitled "Shifting direction of the exposed
image area [ → 174]".
NOTICE! Use points as decimal separators!
4. Continue with the adjustment procedure as described in the chapter
entitled "Automatic adjustment: Pan diaphragm (standard) [ → 188]".
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
menu with the measured values.
D3632.076.01.03.02 09.2016
64 95 258 D3632