Sirona Dental Systems GmbH Table of contents Installation Requirements SINIUS / SINIUS CS / SINIUS TS Table of contents General information....................Notes on the installation prerequisites ............Structure of the document................1.2.1 Identification of danger levels............1.2.2 Formats and symbols used ..............
Page 4
Table of contents Sirona Dental Systems GmbH Installation Requirements SINIUS / SINIUS CS / SINIUS TS SINIUS CS dimensions, scale bar 1:20 ............4.2.1 Dimensions of the treatment room ........... 4.2.2 Side view ..................4.2.3 Top view ................... 4.2.4 Top view with options ...............
Notes on the installation prerequisites This document describes the installation prerequisites for all versions of HYBRID product the SINIUS / SINIUS CS / SINIUS TS treatment center. It contains the following information: ● Required information for practice planning. ● Information for the installer and operator regarding the necessary quality of the air and water supply media.
1 General information Sirona Dental Systems GmbH 1.2 Structure of the document Installation Requirements SINIUS / SINIUS CS / SINIUS TS Structure of the document 1.2.1 Identification of danger levels To prevent personal injury and material damage, please observe the warning and safety information provided in this document. Such...
Sirona accessories or third-party accessories expressly approved by Sirona. The user assumes the risk of using non-approved accessories. If any devices not approved by Sirona are connected, they must comply with the applicable standards, e.g.: ● IEC 60950-1 for information technology equipment (e.g. PC) and ●...
2 Safety information Sirona Dental Systems GmbH 2.4 Power connection Installation Requirements SINIUS / SINIUS CS / SINIUS TS Power connection WARNING Shock hazard It is essential that you switch off the power supply PRIOR to beginning the installation. There is a risk of electric shock. People can be injured or electrical components of the unit destroyed.
Sirona Dental Systems GmbH 3 On-site installation Installation Requirements SINIUS / SINIUS CS / SINIUS TS 3.1 Substrate, floor On-site installation Substrate, floor Unevenness The floor must be level and horizontal in accordance with DIN 18 202. Level floor, steel adapter plate In the case of an uneven floor, the steel adapter plate must be used;...
3 On-site installation Sirona Dental Systems GmbH 3.1 Substrate, floor Installation Requirements SINIUS / SINIUS CS / SINIUS TS 3.1.3 Stability With a concrete floor With a concrete floor/SINIUS ● Use a 12 mm dia. masonry drill to drill the holes. ● The depth of the drill hole (C) in the concrete floor must be at least 80 max.
Sirona Dental Systems GmbH 3 On-site installation Installation Requirements SINIUS / SINIUS CS / SINIUS TS 3.2 Connection to the public drinking water system Connection to the public drinking water system Treatment center isolated from the public drinking water supply Disinfection system, optional...
3 On-site installation Sirona Dental Systems GmbH 3.3 Media quality Installation Requirements SINIUS / SINIUS CS / SINIUS TS Media quality Description of media quality < 0,18 bar > 500 l/min 1 bar = 14.5 psi 5,5 bar < < 7,5 bar 2,5 bar <...
Page 13
Sirona Dental Systems GmbH 3 On-site installation Installation Requirements SINIUS / SINIUS CS / SINIUS TS 3.3 Media quality Water quality Lime deposits and corrosion residues in tap water can lead to the GA231 following malfunctions: ● Premature clogging of the filters in the unit ●...
Page 14
3 On-site installation Sirona Dental Systems GmbH 3.3 Media quality Installation Requirements SINIUS / SINIUS CS / SINIUS TS ● Observe EN 1717 for the protection of public drinking water. Treatment center with disinfection system: The treatment center fulfills the requirements of EN 1717 and that of the DVGW (the German Technical and Scientific Association for Gas and Water).
Installation template We recommend that you order the installation template (REF 33 15 830) from Sirona for laying the pipe ends in the installation field. If necessary, you can also prepare the template yourself based on the diagram below. Table for template 63 22 668 D3561 D3561.021.01.06.02 12.2014...
Page 16
3 On-site installation Sirona Dental Systems GmbH 3.4 Underfloor installation of supply lines Installation Requirements SINIUS / SINIUS CS / SINIUS TS Water supply Pipe 10x1 mm, corner valve outlet 3/8" Compressed air supply Pipe 10x1 mm, corner valve outlet 3/8" Suction line DN40 HT-PP ISO 8283-3 (polypropylene, inner diameter 36.5 mm)
Sirona Dental Systems GmbH 3 On-site installation Installation Requirements SINIUS / SINIUS CS / SINIUS TS 3.4 Underfloor installation of supply lines 3.4.2 Installation of the supply line in the termination panel ✔ An installation template is available or was prepared. Supply lines for Teneo, Sinius 1.
3 On-site installation Sirona Dental Systems GmbH 3.5 Cleaning the air and water pipes Installation Requirements SINIUS / SINIUS CS / SINIUS TS Cleaning the air and water pipes NOTICE Flushing the air and water pipes Chips and other foreign materials could be flushed and/or blown into the treatment center.
Sirona Dental Systems GmbH 3 On-site installation Installation Requirements SINIUS / SINIUS CS / SINIUS TS 3.6 Underfloor installation of the PC connections Underfloor installation of the PC connections Depending on the prevailing local conditions, the existing cable set can be installed in the cable duct of an underfloor installation by an installer.
4 Dimensions, technical data Sirona Dental Systems GmbH 4.1 SINIUS dimensions, scale bar 1:20 Installation Requirements SINIUS / SINIUS CS / SINIUS TS Dimensions, technical data SINIUS dimensions, scale bar 1:20 Dentist element with travel track 4.1.1 Dimensions of the treatment room...
Recommended installation height for wall adapter: 1110 mm (43 3/4“) Cable bushing for control line The control line between SINIUS and the wall adapter are supplied (12.5 mm outer diameter, max. cable length 10 meters (393”). 63 22 668 D3561...
Recommended installation height for wall adapter: 1110 mm (43 3/4“) Cable bushing for control line The control line between SINIUS and the wall adapter are supplied (12.5 mm outer diameter, max. cable length 10 meters (393”). 63 22 668 D3561...
Recommended installation height for wall adapter: 1110 mm (43 3/4“) Cable bushing for control line The control line between SINIUS and the wall adapter are supplied (12.5 mm outer diameter, max. cable length 10 meters (393”). 63 22 668 D3561...
4 Dimensions, technical data Sirona Dental Systems GmbH 4.4 Mounting plates Installation Requirements SINIUS / SINIUS CS / SINIUS TS Mounting plates Adapter plate Adapter plate for Teneo, Sinius If the dental treatment center is being installed as a replacement for an M1 previously installed at this location, an adapter plate is available for this purpose.
Page 39
Sirona Dental Systems GmbH 4 Dimensions, technical data Installation Requirements SINIUS / SINIUS CS / SINIUS TS 4.4 Mounting plates Demonstration chair plate For floors which do not permit permanent connection of the unit (e.g. demo operation at a trade show, floor heating), installation on a steel demo plate is possible.
Information on planning for the practice The following file(s) for practice planning are available for download in the dealer area of the Sirona website under DOWNLOADS => CAAD files: ● PDF file with print symbol - for to-scale printing on paper or adhesive film ●...
Sirona Dental Systems GmbH 4 Dimensions, technical data Installation Requirements SINIUS / SINIUS CS / SINIUS TS 4.6 Technical data Technical data Model designation: SINIUS / SINIUS CS / SINIUS TS Model designation, power ratings Power connection: 100 - 230 V AC ± 10%...
Page 42
4 Dimensions, technical data Sirona Dental Systems GmbH 4.6 Technical data Installation Requirements SINIUS / SINIUS CS / SINIUS TS Transport and storage Temperature: -40 °C – +70 °C Operating and transport conditions 1060 conditions: (-40 °F – 158 °F) Relative humidity: 10% – 95% Air pressure: 500 hPa –...
Page 43
Sirona Dental Systems GmbH 4 Dimensions, technical data Installation Requirements SINIUS / SINIUS CS / SINIUS TS 4.6 Technical data Weight (with packaging and SINIUS Weight, packaging, pressures accessories/without dentist element, assistant element: 63 kg packaging): / 42 kg SINIUS CS dentist element:...
4 Dimensions, technical data Sirona Dental Systems GmbH 4.7 Standards/Approvals Installation Requirements SINIUS / SINIUS CS / SINIUS TS Standards/Approvals ® The treatment center SINIUS complies with the following standards, Compliance with standards, without HF surgery among others: ● IEC 60601-1 (electrical and mechanical safety plus software reliability) ●...
Page 45
Sirona Dental Systems GmbH 4 Dimensions, technical data Installation Requirements SINIUS / SINIUS CS / SINIUS TS 4.7 Standards/Approvals This unit meets the requirements of ATS and may therefore be connected ATS for Australia to the public drinking water supply in Australia.
EMC aspects. SINIUS / SINIUS CS / SINIUS TS complies with the requirements for electromagnetic compatibility (EMC) according to IEC 60601-1-2. SINIUS / SINIUS CS / SINIUS TS is hereinafter referred to as "UNIT". Accessories Making the PC connection The required interface cables can be ordered from Sirona.
Sirona Dental Systems GmbH 5 Electromagnetic compatibility Installation Requirements SINIUS / SINIUS CS / SINIUS TS 5.2 Electromagnetic emission Electromagnetic emission The UNIT is intended for operation in the electromagnetic environment specified below. The customer or user of the UNIT should make sure that it is used in such an environment.
5 Electromagnetic compatibility Sirona Dental Systems GmbH 5.3 Interference immunity Installation Requirements SINIUS / SINIUS CS / SINIUS TS Interference immunity The UNIT is intended for operation in the electromagnetic environment specified below. The customer or user of the UNIT should make sure that it is used in such an environment.
5 Electromagnetic compatibility Sirona Dental Systems GmbH 5.4 Working clearances Installation Requirements SINIUS / SINIUS CS / SINIUS TS Working clearances Recommended working clearances The UNIT is intended for operation in an electromagnetic environment, between portable and mobile HF where radiated RF interference is checked. The customer or the user of...
Sirona Dental Systems GmbH 5 Electromagnetic compatibility Installation Requirements SINIUS / SINIUS CS / SINIUS TS 5.5 Foot control wireless interface Foot control wireless interface Insofar as the treatment center is equipped with a foot control, one wireless module each must be installed in the foot control and in the base of the chair of the treatment center.
We recommend performing an inspection of the circumstances on location at least 4 weeks prior to installation. The checklist should help you when doing this. This can help ensure a smooth procedure on the day that the SINIUS is actually installed. Installation location: ●...
6 Checklist Sirona Dental Systems GmbH 6.3 IT hardware Installation Requirements SINIUS / SINIUS CS / SINIUS TS IT hardware Mode of operation ☑ Mode of operation for Teneo, Sinius ● Stand-alone solution (mini PC required in the base of the chair) ☐...
● Network connection for SINIUS available. ☐ ● Network connection for external PC available. ☐ IMPORTANT! The use of routers between SINIUS and the treatment center PC must be avoided. ● Network configuration plan available. ☐ ● Network jacks have been certified. ☐...