2 Safety instructions
2.5 Electrostatic discharge
16
2.5

Electrostatic discharge

Electrostatic discharge (abbreviated: ESD – ElectroStatic Discharge)
Electrostatic discharge from people can damage electronic components
when the components are touched.
Touch a ground point to discharge static electricity before touching any
boards.
2.6

Switching the unit on

Due to the risk of injury caused by malfunction, no person may be
Safety information for switching on the unit: Service engineer
XGPF-63464; Patient position when switching on
positioned in the unit when it is switched on.
2.7

Condensation

Extreme fluctuations of temperature may cause condensation inside the
Safety information for condensation: Service engineer
unit. Do not switch the unit on before it has reached normal room
temperature. See also the chapter "Technical details"
2.8

Laser light localizer

The system incorporates Class 1 laser products.
Safety information for light localizer: Service engineer
A minimum distance of 10 cm (4") is required between the eye and the
laser. Do not stare into the beam.
Do not use the system with any other lasers, and do not make any
changes to settings or processes that are not described in these
operating instructions. This may lead to a dangerous exposure to
radiation.
2.9

Ventilation slots

Never cover the ventilation slots on the unit under any circumstances,
Ventilation slots
since this may obstruct air circulation. This can cause the unit to overheat.
Under no circumstances may the ventilation slots on the sensor unit be
SL sensor ventilation slots
covered, since otherwise the air circulation in the sensor unit will be
obstructed. This can cause the sensor unit to overheat.
2.10

Qualifications of service personnel

Installation and startup may be carried out only by personnel specifically
authorized by Sirona.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL