Sirona ORTHOPHOS SL 2D Service Manual page 195

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016
A
Adjustment OK
B
C
Adjustment not OK
4. Repeat the procedure starting at Step 2 until the values in the menu's
text boxes are equal to "zero". This means that adjustment was
performed successfully.
IMPORTANT: If automatic adjustment fails to produce values equal
to zero, calculate the adjustment values manually and overwrite the
values in the menu's text boxes with these values [ → 196].
5. If adjustment is OK (the adjustment values in the text boxes are equal
to "zero"), save the values. [ → 175]
The status indicator that prefixes the
element is checked.
6. Remove the needle phantom from the bite block holder of the unit.
7. Continue with the next calibration point. [ → 197]
7 Unit adjustment and calibration
7.2 Adjustment and calibration via the calibration menu
● The shadow of the center needle,
the needle image, and the
auxiliary lines must be coincident
and located one behind the other.
A tolerance (offset of needle from
the central auxiliary line) of ± 0.75
mm is admissible.
● Distance A1 must be 88.6 ± 1 mm.
● Distances A2 must be identical,
each being 44.3 ± 0.5 mm.
● A surrounding white border must
be present.
"Symmetrie adjustment"
195

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orthophos sl 3dOrthophos sl 3d ceph

Table of Contents