Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016
The offset must be determined by reference to an existing exposure.
Based on the default setting, an offset in the range of +15,000 μm to -
30,000 μm can be selected and saved.
1. Call service routine S034.9 [ → 249].
The current offset value is displayed in selection field 1.
2. Select the desired offset value using the arrow keys (A) in selection
field 1 (step width = 500, offset value = 1000 corresponds to a 1 mm
shift of the shadowing limit).
Once the required setting has been selected, the Memory key (B)
lights up.
3. Save the setting [ → 251].
4. Exit the service routine [ → 252].
8.1 Perform service routines via the control panel
8 Service routines