8 Service routines
8.1 Perform service routines via the control panel
254
3. Initiate the radiation.
Radiation is released from any tube assembly position.
There is no rotational movement.
The primary diaphragm is in the last position set.
The radiation time is 14.0 seconds.
IMPORTANT: If you let go of the release button before this
radiation time expires, radiation will be terminated prematurely.
The actual radiation time is
When you release radiation during the cool-down interval, a
countdown of the remaining waiting time is displayed in the title
bar of the control panel (automatic exposure blocking).
Radiation is emitted as long as the release button remains
pressed, but for no longer than 14.0 seconds.
4. Exit the service routine [ → 252].
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
not
displayed.
64 95 258 D3632
D3632.076.01.03.02 09.2016