Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016
Sensor calibration
7.2.2.3
1. In the structure tree, under
element (S010.64).
2. Establish [ → 175] receptivity.
3. Create an exposure (85 kV / 22 mA s) [ → 175].
A message is displayed in the action window to indicate that the
exposure is complete and the status indicator in front of the
"Sensor calibration"
accordingly:
Calibration not OK = status indicator stays red
Calibration OK = status indicator turns green
NOTE: The calibration values are automatically saved.
4. Continue with the next stage of the calibration procedure [ → 230].
7 Unit adjustment and calibration
7.2 Adjustment and calibration via the calibration menu
"3D"
"Sensor calibration"
, click on the
element in the structure tree responds
229