S005: Test Step 4 - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016
During the adjustment procedure, a progress indicator is
displayed in selection field 1.
"CONTINUOUS" is displayed in selection field 2.
After a successful adjustment, the offset value is displayed in
selection field 3. The "Release" note in selection field 1 requests
that you let go of the release button.
3. Exit the service routine [ → 252].
8.1.6.2

S005: Test step 4

Fan test
Selection field
1
* factory settings
1. Call service routine S005.4 [ → 249].
2. Use the arrow keys (A) in selection field 1 to select the code "01" (see
table).
3. Confirm your selection by pressing the R key (B).
The fan starts up.
4. Check the fan for running noise.
5. Exit the service routine [ → 252].
Upon exiting the service routine, the fan is automatically switched off
again.
8.1 Perform service routines via the control panel
Code
Function
00
Fan off*
01
Fan on
8 Service routines