Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016
The exposure is displayed in the exposure window.
4. Repeat the procedure starting at Step 2 until the value in the menu's
text box is equal to "zero". This means that adjustment was
performed successfully.
IMPORTANT: If automatic adjustment fails to produce values equal
to zero, calculate the adjustment values manually and overwrite the
values in the menu's text boxes with these values [ → 220].
5. If adjustment is OK (the adjustment values in the text boxes are equal
to "zero"), save the values. [ → 175]
The status indicator that prefixes the
beam direction"
element is checked.
6. Remove the ceph test phantom from the sensor unit cover.
7. Continue with the next stage of the adjustment procedure. [ → 221]
7 Unit adjustment and calibration
7.2 Adjustment and calibration via the calibration menu
"Adjustment of Ceph main
219