Quality Test Exposure Din 6868-161 (3D) - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
B
A
64 95 258 D3632
D3632.076.01.03.02 09.2016

Quality test exposure DIN 6868-161 (3D)

10.11.3.5
You can use the
volume exposure with a full rotation according to DIN 8686-161. This
function enables you to simulate a quality test, for example, in a similar
way to an acceptance/constancy test.
1. Insert the 3D test phantom into the bite block holder of the unit.
NOTICE! The 3D test phantoms from ORTHOPHOS and GALILEOS
have a similar appearance. Ensure that you use the 3D test phantom
with the "ORTHOPHOS" product designation.
2. Align the 3D test phantom with the aid of the in-built spirit level (B) by
twisting the setting screw (A) horizontally.
3. Press the R key.
The unit moves to its starting position.
4. Call menu
5. In the structure tree, under
"Quality control exposure DIN 6868-161"
In the action area, the menu appears for creating the
control exposure DIN 6868-161"
6. Click on the
Exposure readiness is established. A dialog window displays the
status of readiness to exposure.
The service routine used for the corresponding exposure is
displayed on the control panel, along with the specific exposure
parameters.
"Quality control exposure DIN 6868-161"
"Control exposures"
[ → 428].
"3D control exposures"
"Start acquisition"
button.
10 Maintenance
10.11 Test exposures/Test images
to create a
, select the
element.
"Quality
.
451