Displays And Help Messages During Adjustment And Calibration - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
Help messages
Status messages
64 95 258 D3632
D3632.076.01.03.02 09.2016
7.1.1
Displays and help messages during adjustment and
calibration
The most frequent help and status messages during calibration are listed
below.
H3 01: Move unit to starting position, press the R key.
H3 07: Remove the occlusal bite block.
H3 21: Close the door
H4 02: Plug the sensor into the ceph slot (sensor missing).
H4 03: SIDEXIS 4 is not ready for exposure, make unit ready for
exposure.
H4 04: Plug the sensor into the ceph slot (incorrect sensor plugged in).
"Ready for exposure"
"Exposure not possible"
"Please wait"
"Ready for exposure in XXs"
"X RAY Active!"
"Exposure is performed"
If error message E1 11 20 is displayed on the control panel and/or the
remote control during the calibration process, this does not necessarily
indicate an equipment error. This error message only indicates that the
adjustment or calibration data of the unit is incomplete at this point.
Acknowledge the error message with the R key, if applicable, and
continue the adjustment or calibration procedure.
For assistance with other help messages or error messages displayed
during the adjustment or calibration process, please refer to the section
of these instructions entitled "Messages".
7 Unit adjustment and calibration
7.1 General information on unit adjustment and calibration
163

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orthophos sl 3dOrthophos sl 3d ceph

Table of Contents