4 General operating procedures
4.2 Setting up the X-ray component
"Boot service"
54
Boot service
IMPORTANT
The X-ray component can only be detected via the boot service if the
default network data is set on the unit (see chapter "S037: Test step 2
[ → 322]").
The menu for adding X-ray components opens.
"Boot service"
1. Click the
2. A message window prompts you to switch the new X-ray component
back on (to reboot).
The network is searched for the "booting" X-ray component.
The X-ray components found in the network is displayed with
their network data.

Sirona Dental Systems GmbH

Service Manual ORTHOPHOS SL
button in the structure tree.
D3632.076.01.03.02 09.2016
64 95 258 D3632