9 Repair
9.14 Light barriers
After replacing V1_1, V1_2 or V1_3
After replacing V1_4
After replacing V61_4 or V61_5
After replacing V91_1 or V91_2
394
Cephalometer
● Light barrier for patient diaphragm position detection, ceph: V91_1
● Light barrier for sensor position detection, ceph: V91_2
9.14.2

What has to be done after replacing the light barriers?

➢ Perform a complete unit adjustment or calibration [ → 160].
➢ Perform a function check.
➢ Perform a complete unit adjustment or calibration [ → 160].
1. Perform service routines S034.4 [ → 309] and S034.5 [ → 313].
2. Check the ceph adjustment using the test exposure Ceph - Fixed
point of rotation test exposure [ → 437].
3. Perform a ceph unit adjustment, if necessary [ → 160].
Service Manual ORTHOPHOS SL
V91_2
Sirona Dental Systems GmbH
V91_1
64 95 258 D3632
D3632.076.01.03.02 09.2016