Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016
4.6
Backing up / restoring / installing sensor data
The sensor data can be saved, resaved and installed using the Sirona
Web service.
4.6.1
Access the "Backup / restore sensor data" menu
✔
The unit and computer are turned on.
Open the backup / restore data menu
✔
The unit is logged into the network as an X-ray component.
1. Start an Internet browser such as Internet Explorer or Firefox.
2. In the address line, enter "https://" and the IP address of your unit.
Example: https://192.168.15.240
NOTICE! The IP address of your unit can be found on the Info screen.
3. Confirm your input with the "Enter" key.
A security prompt appears.
4. Click on the link
The Sirona Web service is loaded.
The sub menus of the
tree.
5. Click on the
4.6 Backing up / restoring / installing sensor data
"Continue to this Website"
"User"
menu are displayed in the structure
"Service"
button in the header line of the main menu.
4 General operating procedures
.