Sirona ORTHOPHOS SL 2D Service Manual page 87

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
Error code
Description
E3 06 23
Hardware signal of release button
applied during power-on.
Error code
Description
E5 06 30
Total radiation time exceeded.
Error code
Description
E5 06 31
Partial radiation time exceeded
Error code
Description
E5 06 32
Minimum preheating time not
observed.
Error code
Description
E1 06 40
Tolerance exceeded
VH nom.
64 95 258 D3632
D3632.076.01.03.02 09.2016
Actions required
E3 06 23
● Check cable L5:
1. Switch off the unit.
2. Pull cable L5 off tube assembly.
3. Switch unit on.
4. Perform optical check of L5:
If light is visible ...
● Replace board DX1.
If no light is visible ...
● Replace the tube assembly.
Actions required
E5 06 30
If a CAN bus error had been reported
before ...
● Check the CAN bus.
● Please contact the Sirona Customer
Service Center (CSC) to find out
whether a bug fix by means of a
software update is possible and
perform such an update if necessary.
Actions required
E5 06 31
If a CAN bus error had been reported
before...
● Check the CAN bus.
● Please contact the Sirona Customer
Service Center (CSC) to find out
whether a bug fix by means of a
software update is possible and
perform such an update if necessary.
Actions required
E5 06 32
If a CAN bus error had been reported
before...
● Check the CAN bus.
● Please contact the Sirona Customer
Service Center (CSC) to find out
whether a bug fix by means of a
software update is possible and
perform such an update if necessary.
Actions required
E1 06 40
● Run service routine S005.2.
● Replace the tube assembly.
5 Messages
5.4 List of error messages
see
S. [ → 395]
S. [ → 372]
see
S. [ → 141]
S. [ → 68]
see
S. [ → 141]
S. [ → 68]
see
S. [ → 141]
S. [ → 68]
see
S. [ → 260]
S. [ → 372]
87