S002: Test Step 3 - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

8 Service routines
8.1 Perform service routines via the control panel
256
3. Initiate the radiation.
Radiation is released from any tube assembly position.
There is no rotational movement.
The primary diaphragm is in the last position set.
The radiation time corresponds to the selected radiation time.
Important: If you let go of the release button before this radiation
time expires, radiation will be terminated prematurely. The actual
radiation time is not displayed.
When you release radiation during the cool-down interval, a
countdown of the remaining waiting time is displayed in the title
bar of the control panel (automatic exposure blocking).
4. Exit the service routine [ → 252].
8.1.5.2

S002: Test step 3

Radiation without rotary movement; selectable kV/mA level and
maximum radiation time; primary diaphragm fully opened with respect
to PAN position.
WARNING
Unit is radiating X-rays
Excess exposure to X-rays is detrimental to health.
➢ Use the prescribed accessories for radiation protection.
➢ Do not stay in the X-ray room during exposure. Move as far away
from the unit as the coiled cable for the release button allows you to.
Selection field
1
2
Factory settings: 60kV/8mA, 2s
1. Call service routine S002.3 [ → 249].
Calling this service routine moves the primary diaphragm to the
new position (fully opened with respect to PAN position). During
this time, a progress indicator is displayed in selection field 1.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
Parameters
Range of values
kV/mA level
60 kV/3mA –
90kV/12mA –
Radiation time
0.1s – 4s
D3632.076.01.03.02 09.2016