Service Manual ORTHOPHOS SL
move the slide
C
F
E
Locking the slide
64 95 258 D3632
D3632.076.01.03.02 09.2016
9.2.1.1.1
Moving the slide with the "height adjustment" service kit,
REF. 62 57 518
1. Insert the clamp through the opening (A) in the stand and rotate it 90°.
Tighten the nut securely.
2. Loosen the two screws (C) and remove the cover (D).
C
D
3. Remove the 1st screw (E) on the spindle holder (F).
4. Insert the locking pin (B) into this opening.
5. Now remove the 2nd screw (E) on the spindle holder (F).
6. Attach the socket wrench (SW19) to the spindle. Remove the locking
pin and then turn the slide up to the desired height using the socket
wrench.
Clockwise rotation of spindle = slide moves up
Counterclockwise rotation of spindle = slide moves down
7. Reinsert the locking pin.
➢ Now move the clamp (A) directly underneath the slide.
The slide is now locked in this position for further repair work.
9.2 Height adjustment motor (M1_4)/spindle
9 Repair