S034: Test step 8
8.1.20.4
User-specific determination of the shadowing limit for C1 and C2
For ceph programs C1 and C2, the default setting of the shadowing limit
can be assigned an individual offset. In this way the shadowing can be
shifted within certain limits.
C1 / C2
A
Shadowing limit
The offset must be determined by reference to an existing exposure.
Based on the default setting, an offset in the range of +30,000 μm to -
15,000 μm can be selected and saved.
1. Call service routine S034.8 [ → 249].
The current offset value is displayed in selection field 1.
8.1 Perform service routines via the control panel
m
+30.000 m
m
-15.000 m