Quality test exposure (3D)

10.11.3.4
You can use the
exposure with a full rotation. This function enables you to simulate a
quality test, for example, in a similar way to an acceptance/constancy
test.
1. Insert the constancy test phantom (A) into the bite block of the unit.
2. Call menu
3. In the structure tree, under
"Quality control exposure Sirona"
In the action area, the menu appears for creating the
control exposure Sirona"
4. Click on the
Exposure readiness is established. A dialog window displays the
status of readiness to exposure.
The service routine used for the corresponding exposure is
displayed on the control panel, along with the specific exposure
parameters.
"Quality control exposure Sirona"
"Control exposures"
[ → 428].
"3D control exposures"
element.
.
"Start acquisition"
button.