7 Unit adjustment and calibration
7.2 Adjustment and calibration via the calibration menu
226
7.2.2
3D adjustment/calibration
IMPORTANT
For the 3D adjustment and calibration, the values in the input fields do
not have to be equal to zero.
Confirmation of a successful adjustment and calibration is the message
in the calibration menu action window; e.g. "Adjustment successful",
"Values can be adopted", etc.
Sensor adjustment
7.2.2.1
1. Insert the 3D calibration block into the bite block holder on the unit.
[ → 179]
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016