Safety instructions

2.1

Potentially lethal shock hazard

DANGER
XGPF-36475; Risk of electric shock
Perilous shock hazard!
When replacing parts or in the event of extensions in the power supply
line, power switch, board DX32, or X-ray tube assembly area, there is
danger to life due to a potentially lethal shock hazard.
Before you start:
➢ It is essential to ensure the unit is not supplied with voltage.
➢ Always disconnect the unit from the junction box of the building
installation.
2.2

Modifications to the unit

Modifications to this unit which might affect the safety of the system
owner, patients or other persons are prohibited by law!
For reasons of product safety, this product may be operated only with
original Sirona accessories or third-party accessories expressly approved
by Sirona. The user is responsible for any damage resulting from the use
of non-approved accessories.
2.3

Fixed connection

DANGER
Potentially lethal shock hazard!
Fixed connection!
Installing a mains plug instead of the specified fixed connection infringes
international medical regulatory actions and is prohibited. In case of
error, this puts patients, users, and other parties seriously at risk.
2.4

Electromagnetic compatibility

The unit complies with the requirements of standard IEC 60601-1-2.
Medical electrical equipment is subject to special EMC-related
precautions. It must be installed and operated as specified in the
document "Installation Requirements".
If high-voltage systems, radio link systems or MRI systems are located
within 5 m of the unit, please observe the specifications stated in the
installation requirements.
Portable and mobile RF communications equipment may affect medical
electrical equipment. Therefore, the use of mobile wireless phones in
medical office or hospital environments must be prohibited.
2 Safety instructions