7 Unit adjustment and calibration
7.2 Adjustment and calibration via the calibration menu
184
6. If adjustment is OK (the adjustment values in the text boxes are equal
to zero), save the values. [ → 175]
The status indicator that prefixes the
element is checked.
7. Remove the needle phantom from the bite block holder of the unit.
8. Continue with the next stage of the adjustment procedure. [ → 187]
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
"Sensor adjustment"
64 95 258 D3632
D3632.076.01.03.02 09.2016