Ceph - Secondary Diaphragm (2D) - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016

Ceph - secondary diaphragm (2D)

10.11.2.6
You can use the
adjustment of the Ceph secondary diaphragm.
1. Call up the menu
2. Select the
tree under
In the action area, the menu appears to create the exposure for
"Ceph - secondary diaphragm"
the
3. Click on the
Exposure readiness is established. A dialog window displays the
status of readiness to exposure.
The service routine used for the corresponding exposure is
displayed on the control panel, along with the specific exposure
parameters.
4. Press the release button. Press and hold down the button until the
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal that indicates the end of the
exposure (double beep) sounds (if configured).
The exposure is displayed in the exposure window.
For notes on evaluating the exposure, refer to the section "Unit
adjustment and calibration [ → 160]"/"Adjustment of the ceph
secondary diaphragm [ → 204]".
"Ceph - secondary diaphragm"
"Control exposures"
"Ceph - secondary diaphragm"
"2D control exposures"
.
"Start acquisition"