Location 20: Sensor Unit, General - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
Error code
Description
E6 20 01
Malfunction when starting the
sensor unit.
Configuration does not match unit or
communication problems following
unit start.
Error code
Description
E6 20 03
Communication fault for sensor unit. ● Check plug connections of fiber-optic
Error code
Description
E6 20 06
Communication fault for sensor unit. ● Check plug connections of fiber-optic
Error code
Description
E6 20 07
Communication fault for sensor unit. Check plug connections of fiber-optic cables to
64 95 258 D3632
D3632.076.01.03.02 09.2016
5.4.9

Location 20: Sensor unit, general

Actions required
E6 20 01
● Check unit configuration (sensors) via
service routine S017.2.
● Restart the unit:
1. Switch off the unit.
2. Wait 1 minute.
3. Switch unit on.
4. Repeat procedure and observe causal
error messages.
If the error occurs repeatedly ...
● Check cable and plug connections on
board DX831.
Actions required
E6 20 03
cables to boards DX83 (X800) and DX1
(J901).
Actions required
E6 20 06
cables to boards DX83 (X800) and DX1
(J901).
● Restart the unit:
1. Switch off the unit.
2. Wait 1 minute.
3. Switch on the unit.
4. Repeat procedure.
If the error occurs repeatedly ...
● Please contact the Sirona Customer
Service Center (CSC) to find out whether
a bug fix by means of a firmware update is
possible and perform such an update if
necessary.
Actions required
E6 20 07
boards DX83 (X800) and DX1 (J901).
5 Messages
5.4 List of error messages
see
S. [ → 278]
S. [ → 153]
see
S. [ → 30]
see
S. [ → 30]
S. [ → 68]
see
S. [ → 30]
113