Adjusting The Ms Laser Light Localizer (Pan) - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

9 Repair
9.7 Laser light localizers
A
360

Adjusting the MS laser light localizer (PAN)

9.7.3.2
The "Front slide" cover must have been removed [ → 37].
1. Insert the forehead and temple supports (see also Operating
Instructions).
2. Switch the unit on via the switch (A) (see also Operating Instructions).
B
C
3. Touch the PAN symbol at the top of the touchscreen.
4. Touch the R key on the touchscreen.
5. Press the light localizer key on the control panel.
The X-ray radiation indicator (B) lights up briefly.
After approx. 2 seconds, the green LED (C) in the upper part of
the control panel lights up. This LED remains lit as long as the
unit is on.
The start screen appears on the touchscreen of the Easypad and
the system's self-adjustment routine starts running (for approx. 1
minute). The diaphragm moves into position. Forehead and
temple supports on the panoramic unit are moved to the home
position.
On completion of the self-adjustment routine, the main menu
appears on the touchscreen. Help message H301 prompts you to
move the unit into the starting position.
The program group PAN is selected.
The unit moves to its starting position for PAN exposures.
The light localizers are switched on.
The vertical laser beam must be displayed in the center of the
forehead support and the bite block holder.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orthophos sl 3dOrthophos sl 3d ceph

Table of Contents