Sirona ORTHOPHOS SL 2D Service Manual page 213

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016
3. Create an exposure (80kV / 14 mA; 0.3 s) [ → 175].
Adjustment value S1 is calculated automatically from the
exposure and entered in the text box of the
secondary diaphragm (Quickshot)"
The exposure is displayed in the exposure window.
7 Unit adjustment and calibration
7.2 Adjustment and calibration via the calibration menu
menu.
"Adjustment of Ceph
213