Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016

Dosimetry DIN 6868-161 (3D)

10.11.3.8
A dosimeter for pulsed radiation (e.g. Mult-O-Meter 512L) is required for
dosimetry.
You can use the
dosimetry according to DIN 6868-161 independently of the unit
adjustment or calibration.
1. Attach the Mult-O-Meter sensor in the middle of the sensor unit.
2. Select a program from the 3D program group on the control panel.
The unit moves to the 3D exposure position.
3. Call up the
4. Select the
structure tree under
In the action area, the menu appears to create the
measurement DIN 6868-161"
5. Click on the
Exposure readiness is established. A dialog window displays the
status of readiness to exposure.
The service routine used for the corresponding exposure is
displayed on the control panel, along with the specific exposure
parameters.
"Dose measurement DIN 6868-161"
"Control exposures"
menu [ → 428].
"Dose measurement DIN 6868-161"
"3D control exposures"
.
"Start acquisition"
button.
10 Maintenance
10.11 Test exposures/Test images
menu to perform
element in the
.
"Dose
455