Checking The Laser Light Localizers - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016
10.8

Checking the laser light localizers

CAUTION
Risk of injury to eyes.
The unit contains lasers of Class 1.
Keep a distance of at least 4" (10 cm) between eye and laser. Do not
look into the laser beam.
Checking the laser light localizers
Preparing the test
1. Insert the forehead and temple supports.
2. Set the main switch (A) to I (see also Operating Instructions).
3. Wait for approx. 1 minute.
4. Touch the R key on the touchscreen.
The unit moves to its starting position.
10 Maintenance
10.8 Checking the laser light localizers
417