5. Loosen the two rear screws (C) on the tube assembly.
6. CAUTION! The tube assembly is heavy!
Hold the tube assembly firmly in place, loosen the two front screws D
(3-4 turns) and remove the tube assembly toward the front.
Tip: If you leave the two front screws on the rotating element, you can
immediately hang the tube assembly on them when reinstalling it.
7. Hang the new tube assembly on the two front screws of the rotating
element and tighten them securely.
8. Insert the two rear screws and tighten them firmly.
9. Plug cables L3, L5 and L6 as well as the ground cable back onto
board DX6 and reattach the cables to the rubber grommets.
10. Reattach the cover plate.
11. Reassemble the diaphragm unit [ → 370].
12. Reattach the covers "Front tube assembly" and "Rear tube assembly"
[ → 37].
9 Repair
9.11 X-ray tube unit