Enabling Exposure Readiness; Taking An Exposure; Save Values - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016
7.1.3

Enabling exposure readiness

To take an exposure, the system must first be made ready for exposure.
The calibration menu must be opened [ → 164].
Select the corresponding element in the structure tree.
"Start acquisition"
➢ Click the
Exposure readiness will be established. A dialog window displays
the status of readiness to exposure.
The service routine used for the corresponding exposure is
displayed on the control panel, along with the specific exposure
parameters.
7.1.4

Taking an exposure

The calibration menu must be opened.
Select the corresponding element in the structure tree.
The unit must be made ready for exposure [ → 175].
➢ Press the release button. Press and hold down the button until the
exposure is complete, the X-ray image is displayed in the exposure
window, and the acoustic signal indicating the end of the exposure
(double beep) sounds (if configured [ → 287]).
7.1.5

Save values

The adjustment and calibration must have been performed correctly.
➢ To save the adjustment and calibration values, click the
save"
button in the calibration menu.
The adjustment and calibration is saved.
The saved adjustment and calibration is identified in the structure
tree by a check mark (adjustment) or a green traffic light
(calibration).
7 Unit adjustment and calibration
7.1 General information on unit adjustment and calibration
button in the calibration menu.
"Validate and
175