8 Service routines
8.1 Perform service routines via the control panel
270
The kV/mA level is displayed in selection field 1.
The corresponding radiation time is displayed in selection field 2.
3. Use the arrow keys (A) to select the required kV/mA level, e.g. "85kV/
5mA".
The corresponding radiation time "3.2 s" is displayed in selection
field 2.
NOTE: The radiation time cannot be selected separately.
Once the required setting has been selected, the Memory key (B)
lights up.
4. Save the setting.
5. Initiate the radiation.
Continuous radiation of 85kV/5mA is applied for 3.2 s.
IMPORTANT: If you let go of the release button before the
maximum radiation time has elapsed, radiation is terminated
prematurely and the exposure is interrupted.
When you release radiation during the cool-down interval, a
countdown of the remaining waiting time is displayed in the
Easypad title bar (automatic exposure blocking).
6. Read the dose shown on the Mult-O-Meter.
7. Exit the service routine.
8. Remove the Mult-O-Meter from the sensor unit again.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632