Sirona Dental Systems GmbH

Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016
4

General operating procedures

4.1

Switching the unit on

WARNING
X-rays
Be sure to observe the radiation protection regulations applicable in
your country.
➢ No person may be positioned in the unit when it is switched on.
NOTICE
Damage to the unit
Check the room height before you raise the unit.
➢ If the room height is less than 2.27 m (89 3/8") or 2.30 m (90 1/2")
for installation with the floor stand, you must limit the maximum
travel height.
NOTICE
Damage to the unit
Make sure that no objects are in the movement range of the device
when it moves (exception: Calibration block and positioning aids).
NOTICE
Fluctuations in temperature can cause condensation to form in the unit.
Electrical components are destroyed by short circuits.
➢ Do not switch the unit on until the temperature of the unit has
adapted to the ambient temperature and the condensation has
evaporated.
Also note the operating temperature specified for this device.
NOTICE
Waiting time when switching on and off
The unit must not be switched on/off constantly.
Constant switching on and off reduces the service life of individual unit
components and results in increased power consumption.
➢ After switching the unit off, wait for approx. 60 seconds before
switching it on again.
NOTICE
The surface of the touchscreen is sensitive.
The touchscreen can be damaged or its surface scratched.
➢ Never use pointed objects such as ballpoint pens, pencils, etc. to
operate the touchscreen.
➢ Only use your fingertips to operate the touchscreen.
4 General operating procedures
4.1 Switching the unit on
47