Pan - dark current exposure (2D)

10.11.2.5
You can use the
sensor and the data path.
1. Call up the menu
2. Select the
"2D control exposures"
under
The menu for creating the
in the action window.
3. Click on the
Exposure readiness is established. A dialog window displays the
status of readiness to exposure.
The service routine used for the corresponding exposure is
displayed on the control panel, along with the specific exposure
parameters.
4. Press the release button. Press and hold down the button until the
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal that indicates the end of the
exposure (double beep) sounds (if configured).
The exposure is displayed in the exposure window.
See below for reference image.
"Pan - dark current exposure"
"Control exposures"
"Pan - dark current exposure"
.
"Pan - dark current exposure"
"Start acquisition"
button.
10 Maintenance
10.11 Test exposures/Test images
menu to check the DCS
[ → 428].
element in the structure tree
appears