14. Remove the motor-side end piece from the cable track.
15. With defective cable L117: Unscrew cable L117 from the interface
board and remove the shield.
If cable L117 should be intact and used again, this step is not
necessary. Unless it is not possible to lay down the cable track flat
near the stand (see next step).
16. Remove the cable ties from the cable track and lay the cable track
down on a flat surface stretched out.
17. CAUTION! You must observe the position of connector X303 of cable
L108 (see image).
Carefully pull both cables (together) out of the cable track and the
fabric tube.
18. With defective cable L108: Wrap the connector X303 of the new cable
L108 with adhesive tape to protect the detent against breaking off.
19. Lay the cable track down on a flat surface stretched out.
20. Fasten the two (new) cables together with adhesive tape above the
flag labels.