Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
Image LED V501
Pan LED V500
64 95 258 D3632
D3632.076.01.03.02 09.2016
S032: Test step 50
8.1.18.2
Self-test of the FP image data path
1. Call service routine S032.50 [ → 249].
Once the service routine has been selected, "FP test" is
displayed in selection field 1.
2. Touch the T key (B).
The FP image data path is tested.
If the image data path is OK, "Success" is displayed in the
selection field 1.
3. Exit the service routine [ → 252].
8.1.18.3
Explanations on the test procedure
The synchronized readout sequence of the sensor is performed with real
data transmissions and CAN bus commands.
The image data are analyzed in the memory of the DX11.
It is possible to evaluate the LEDs on DX1 (image, pan) during the
synchronized readout sequence.
LED is lit
LED not lit
LED is lit
LED not lit
LED not lit
In addition, "wiggle tests" of cables or ring rotation tests can also be
performed.
8.1 Perform service routines via the control panel
Synchronized readout sequence
active
Standby mode
Sensor unit available / standby mode
Sensor unit not available / standby mode
Sensor unit available / synchronized readout
sequence active
8 Service routines
305