7 Unit adjustment and calibration
7.2 Adjustment and calibration via the calibration menu
230

Diaphragm configuration

7.2.2.4
The 3D calibration block
1. In the structure tree, under
"5x5"
the
element (S030.11).
"5x5"
The
2. Establish [ → 175] receptivity.
3. Create an exposure (85 kV / 0.34 mAs) [ → 175].
A message is displayed in the action window to indicate whether
adjustment has been successful and the status indicator in front
"5x5"
of the
Calibration not OK = status indicator stays red
Calibration OK = status indicator turns green
NOTE: The calibration values are automatically saved.
4. Click on the
tree (S030.12).
"8x8"
The
5. Establish [ → 175] receptivity.
is not
in the bite block holder on the unit.
"3D"
"Diaphragm calibration"
/
volume is selected.
element in the structure tree responds accordingly:
"Next"
button or select the volume
volume is selected.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
, click on
"8x8"
in the structure
64 95 258 D3632
D3632.076.01.03.02 09.2016