7 Unit adjustment and calibration
7.2 Adjustment and calibration via the calibration menu
230
Diaphragm configuration
7.2.2.4
✔
The 3D calibration block
1. In the structure tree, under
"5x5"
the
element (S030.11).
"5x5"
The
2. Establish [ → 175] receptivity.
3. Create an exposure (85 kV / 0.34 mAs) [ → 175].
A message is displayed in the action window to indicate whether
adjustment has been successful and the status indicator in front
"5x5"
of the
Calibration not OK = status indicator stays red
Calibration OK = status indicator turns green
NOTE: The calibration values are automatically saved.
4. Click on the
tree (S030.12).
"8x8"
The
5. Establish [ → 175] receptivity.
is not
in the bite block holder on the unit.
"3D"
"Diaphragm calibration"
/
volume is selected.
element in the structure tree responds accordingly:
"Next"
button or select the volume
volume is selected.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
, click on
"8x8"
in the structure
64 95 258 D3632
D3632.076.01.03.02 09.2016