Adjustment And Calibration Via The Calibration Menu; 2D Adjustment/Calibration; Pan Sensor Adjustment - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
Only for Germany
64 95 258 D3632
D3632.076.01.03.02 09.2016
7.2
Adjustment and calibration via the calibration
menu
IMPORTANT
After every adjustment and calibration of the unit, the reference values
for the constancy measurement must be recalculated and entered in the
"Test results" form, "Reference value" column (see technical document
"ORTHOPHOS SL 2D/3D Acceptance test DIN 6868 Part 151 / Part
161"
(REF 65 61 059).
7.2.1

2D adjustment/calibration

Pan sensor adjustment

7.2.1.1
7 Unit adjustment and calibration
7.2 Adjustment and calibration via the calibration menu
181