Head Support; Replacing The Headrest - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
Removing the defective headrest
Installing the new headrest
Aligning the new headrest
64 95 258 D3632
D3632.076.01.03.02 09.2016
9.5

Head support

9.5.1

Replacing the headrest

1. Hold the headrest firmly from below, loosen screw (A), and remove
the defective headrest.
Tip: To remove the headrest, this must be tilted slightly.
2. Pull cable L18 off connector X1 on board DX5 [ → 30].
1. Plug cable L18 into connector X1 on board DX5 of the new headrest
[ → 348].
2. Position the new headrest in the unit and screw it on loosely with the
screw (A) until a slight stop is perceptible.
1. Switch the unit on.
2. Switch the light localizers on and align the headrest so that the MS
light beam strikes the center of the forehead support.
3. IMPORTANT: Make sure that the headrest does not turn when you
tighten the screw. Tighten the screw (A).
9.5 Head support
A
9 Repair
347

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orthophos sl 3dOrthophos sl 3d ceph

Table of Contents