8 Service routines
8.1 Perform service routines via the control panel
318
2. Select the desired offset value using the arrow keys (A) in selection
field 1 (step width = 500, offset value = 1000 corresponds to a 1 mm
shift of the shadowing limit).
Once the required setting has been selected, the Memory key (B)
lights up.
3. Save the setting [ → 251].
4. Exit the service routine [ → 252].
S034: Test step 9
8.1.20.5
User-specific determination of the shadowing limit for C3 and C3 F
For ceph programs C3 and C3F, the default setting of the shadowing limit
can be assigned an individual offset. In this way the shadowing can be
shifted within certain limits.
C3 / C3F
A
Shadowing limit
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
m
+15.000 m
m
-30.000 m
D3632.076.01.03.02 09.2016
A