IMPORTANT
After replacing boards or modules containing boards, check to make
sure that the software version of the module corresponds to the current
software status of the system. The software versions of the modules can
be queried by running service routine S008.2 [ → 267] or using the "Unit
information" [ → 59]. You can also check the info screen in advance to
determine whether the current software constellation is permissible. If
this is not the case, the version number of the entire software is labeled
with an asterisk (e.g. V05.08.02*).
IMPORTANT
When replacing modules, be sure to note which ones contain boards
and follow the instructions in the section titled Measures following
replacement of boards. Also check whether the current software CD or
the SIRONA dealer page contains any additional more up-to-date
information about module replacement.
Be sure to follow the instructions about how to proceed following module
replacement. You will find this information at the end of each set of
repair instructions.
IMPORTANT
Clean the device after it has been serviced. To do this, use only mild
commercially available cleaning agents (see
Operating Instructions"
9.1

Safety checks

Once repairs are completed, the circuit breaker test and unit leakage
Safety checks
current test must be carried out (see chapter "Checking the circuit
breaker" and "Checking the unit leakage current").
"ORTHOPHOS SL
(REF 64 94 947).
9 Repair
9.1 Safety checks