Ceph - Test Image (2D) - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016

Ceph - test image (2D)

10.11.2.8
You can use the
the unit and the operability of the Ceph sensor.
1. Call up the menu
2. Select the
"2D control exposures"
The menu for creating the test image appears in the action area.
3. Click on the
Exposure readiness is established. A dialog window displays the
status of readiness to exposure.
The service routine used for the corresponding exposure is
displayed on the control panel, along with the specific exposure
parameters.
4. Press the release button. Press and hold down the button until the
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal that indicates the end of the
exposure (double beep) sounds (if configured).
The exposure is displayed in the exposure window.
See below for reference image.
"Ceph - test image"
menu to check the 2D data path of
"Control exposures"
"Ceph - test image"
element in the structure tree under
.
"Start acquisition"
button.
10 Maintenance
10.11 Test exposures/Test images
[ → 428].
441