Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
After resetting the values
64 95 258 D3632
D3632.076.01.03.02 09.2016
7.2.3
Resetting adjustment/calibration
In an absolute emergency, the settings made for adjustment and
calibration can be reset in full or in part to the factory settings and/or
modified manually in the
NOTICE
Create a data backup of the adjustment/calibration data in all cases
before you reset the adjustment and calibration.
1. To reset adjustment and calibration data, click in the structure tree
"Reset"
under
3D"
(depending on what data you would like to reset).
2. To start resetting the adjustment and calibration data, click on the
"Reset"
button.
IMPORTANT
"Reset 3D"
With the
"Diaphragm adjustment"
"Sensor calibration"
from the
calibration"
menus are retained.
If the adjustment and calibration has been reset, the unit must be
readjusted and recalibrated.
7 Unit adjustment and calibration
7.2 Adjustment and calibration via the calibration menu
"Adjustment/calibration"
"Reset Pan"
on the element
"Sensor adjustment"
function, only the
adjustments are reset. The calibration values
"Diaphragm calibration"
,
menu.
"Reset Ceph"
"Reset
,
or
and
"Geometry
and
235